presents
Wheeless' Textbook of Orthopaedics
www.wmt.com
Tracking Pixel

Fixation of intertrochanteric fractures of the femur. A randomised


prospective comparison of the gamma nail and the dynamic hip screw. AU: Bridle-SH; Patel-AD; Bircher-M; Calvert-PT SO: J-Bone-Joint-Surg-Br. 1991 Mar; 73(2): 330-4 AB: We have prospectively compared the fixation of 100 intertrochanteric fractures of the proximal femur in elderly patients with random use of either a Dynamic Hip Screw (DHS) or a new intramedullary device, the Gamma nail. We found no difference in operating time, blood loss, wound complications, stay in hospital, place of eventual discharge, or the patients' mobility at final review. There was no difference in failure of proximal fixation: cut-out occurred in three cases with the DHS, and twice with the Gamma nail. However, in four cases fracture of the femur occurred close to the Gamma nail, requiring further major surgery. In the absence of these complications, union was seen by six months in both groups.



Original Text by Clifford R. Wheeless, III, MD.